45
3. HART Device Output
Subject
Type
Valid Value
Comment
PV / Range Unit
Editable
Select from 25 default units
or from 5 custom units
HART device pressure measurement
unit, see section 3.1.2
PV LRV
Editable
Within HART device sensor
range
Process variable lower range
PV URV
Editable
Within HART device sensor
range
Process variable upper range
Transfer function
Editable
Linear / Square Root
Transfer function type
Alarm State
Read Only
N/A
Alarm State
Damping
Editable
1~100
Damping time, only support one digit
after decimal point
Polling Address
Editable
0~15
HART device polling address
Burst Mode
Editable
On / Off
Burst mode status
Burst Command
Editable
1 / 2 / 3
Burst command, depends on HART
device
Table 16 HART Device Output