![Additel ADT760 User Manual Download Page 81](http://html1.mh-extra.com/html/additel/adt760/adt760_user-manual_2848889081.webp)
69
Select the calibration point which needs to be changed, then input value
◆
3 points for current output calibration
◆
Default current output calibration points are 1mA, 10mA, 20mA
◆
Additel does not recommend changing calibration points.
3. Calibration Execution:
1). Press
to start calibration process
2). Output current from 760 calibrator to standard
3). When standard measuring value is stable, press correspondent
Measured Value
and input
measuring value displayed on
the standard then select Enter button or
to confirm and then
press
to record data and move to the next calibration point.
5.5.1.3 Internal Pressure Module Calibration
1. Output Connection:
Select Internal / External pressurizing;
◆
For Internal pressurizing: Please connect 760 calibrator pressure outlet port
to the standard
ensuring the standard is in measurement mode
◆
For External pressurizing: Please connect 760 calibrator pressure outlet port
with the standard,
ensuring the standard is in pressure generation mode
2. Calibration point setup:
Select the calibration point which needs to be changed, then input value
◆
3 points for internal pressure module calibration
◆
Default calibration points are the lower limit, zero point, and the upper limit of internal pressure
www.
.com
1.800.561.8187
Summary of Contents for ADT760
Page 1: ...www com information itm com 1 800 561 8187...
Page 2: ...I www com information itm com 1 800 561 8187...
Page 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Page 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Page 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Page 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Page 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Page 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...