70
module
◆
Additel does not recommend changing calibration points
NOTE: The lower limit of some internal pressure modules may exceed the lower limit of 760
calibrator pump, please use External pressurizing to calibrate these internal pressure
modules.
3. Calibration Execution:
1). Press
to start calibration process
2). Pressurization:
①
Internal pressurizing:
Generate pressure using the 760 internal pump. When the pressure is stable press Reference
value and input the value and select the Enter key or
to confirm, then press
to record
data and move to the next calibration point.
②
External pressurizing:
Output pressure from standard to 760 calibrator, and not the standard output value shall be the
same as correspondent calibration
set point. When Measured Value is stable, press
to record
data and move to the next calibration point.
5.5.1.4 Auto Tune
This function is only used during when recommended by Additel Corporation to service the calibrator
5.5.1.5 Pressure Sensor
This function is for positive and negative pressure sensor check with atmosphere.
www.
.com
1.800.561.8187
Summary of Contents for ADT760
Page 1: ...www com information itm com 1 800 561 8187...
Page 2: ...I www com information itm com 1 800 561 8187...
Page 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Page 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Page 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Page 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Page 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Page 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...