27
3.1.3 Pressure Output
1. Pressure Output Control Settings
:
Press on the Pressure Window to enter the control settings menu:
Subject
Valid Value
Comment
Pressure Type
Diff. / Gauge / Absolute The pressure type displayed will depend on the ADT760 model
Resolution
4 / 5 / 6
Display digit
Slew Rate
0~100%
Pressurization speed, depends on internal pressure module
Stability Limit
±
(
0.003~1
)
% FS
Pressure control stability condition
Vent Limit
Depends on 760
model
Pressure at which the vent valves will open to atmosphere
Auto Zero
On / Off
Pressure measurement will auto zero every time the unit is
vented (not available with all models)
Set Point Limits
On / Off
Limits the pressure output ranges
Set Point Limits (on)
Depends on internal
pressure module
User editable for 0% and 100% of span ranges
Module Information
N/A
Information is read only and includes range, pressure type,
accuracy, SN, version, and manufacture date
Running Information
N/A
Information is read only and includes supply pressure, vacuum
pressure, and atmosphere
Table 9 Pressure Output Control Settings
www.
.com
1.800.561.8187
Summary of Contents for ADT760
Page 1: ...www com information itm com 1 800 561 8187...
Page 2: ...I www com information itm com 1 800 561 8187...
Page 12: ...XI Figure 19 REF and Outlet Port Filter 98 www com information itm com 1 800 561 8187...
Page 18: ...6 1 2 Basic Structure Figure 1 Basic Structure www com information itm com 1 800 561 8187...
Page 28: ...16 Figure 2 Standard Equipment www com information itm com 1 800 561 8187...
Page 40: ...28 2 Connection Figure 8 Output Connection www com information itm com 1 800 561 8187...
Page 148: ...136 Step 4 Click on Manual Install www com information itm com 1 800 561 8187...
Page 158: ...146 Step 2 Select No to skip update process www com information itm com 1 800 561 8187...