30
③
Auto Step
:
Press
on Pressure Window, then select Auto Step
:
1) Auto Step settings:
Subject
Valid Value
Comment
Range
Depends on internal pressure module
The start and end points of auto step
Step Mode
By
Points/
Percent /
Size
Calculation
method
for
pressure
change of each step
Points
Pressure change of each step
= Range ÷ Points
Pressure change calculation
Percent
Pressure change of each
= Range × Percent ÷ 100
Step Size
Pressure change of each
= Step Size
Stroke
Round Trip / One Way
Auto Step stroke type
Dwell Time
0~100,000,000
(
s
)
Retention time after each step
Repeat
1 /
+∞
Repeating criteria
Table 10 Auto Step
2) Press , then select Start.
4. Discontinue Control or Pressurization:
◆
Manual Input and Manual Step:
Press Measure to pause and press Control to restart.