20
2.2 Internal Pressure Module
The Additel 760 Calibrator comes with an internal pressure module which covers the full span of the
unit. This pressure module is user-switchable and Additel offers a large variety of different range
pressure modules.
◆
To switch the internal pressure module:
1)VENT the system pressure (see section 5.5.3)
2)Turn off the calibrator, disconnect the power adapter if it is connected, and remove the battery
3)Unscrew both screws on the pressure module until the screw pin is loose and the screw pin can
be extended out
4)Firmly pull the sensor out of the calibrator
5)Insert a new internal pressure module (please pay attention to module direction)
6)Tighten both screws on the pressure module until the screws reach the bottom
Figure 4 Internal Pressure Module Switching