7352 Series Digital Multimeter Operation Manual
7.1.5 AC Current Measurement (ACI, ACI(AC+DC))
7-9
Additional error depending on the crest factor (For a non sine-wave input current)
(% of r % of range)
Additional error to the ACI(AC+DC)-Ach measurement
Range
Measurement accuracy *1
(% of r digits)
20 Hz - 45 Hz
45 Hz - 1 kHz
1 kHz - 5 kHz
200
A
0.4+20
0.3+20
0.3+10
2000
A
0.5+20
0.35+10
0.35+20
20 mA
0.4+20
0.3+20
0.3+10
200 mA
0.4+20
0.3+20
0.3+10
2000 mA
0.5+20
0.35+10
0.7+20
*1:
For sine-wave input When RATE is set to FAST or MED, the digit term is reduced
to 1/10 of its specification.
Range
Crest factor
1 - 2
2 - 3
200
A
0+0.05
0+0.15
2000
A
0+0.05
0+0.15
20 mA
0+0.05
0+0.15
200 mA
0+0.05
0.1+0.15
2000 mA
0+0.05
0.1+0.15
FAST
MED
(% of reading)
20 - 45 Hz
1.24%
0.72%
45 Hz - 1 kHz
0.06%
–
Maximum allowable
applied current
Input protection
Fuse replacement
Between mA and
COM A terminals
2 A 2 A/250 V fast-blow fuse which is com-
pliant with IEC60127 sheet1
On the rear panel