7352 Series Digital Multimeter Operation Manual
7.1.2 AC Voltage Measurement
7-4
ACV(AC+DC)-Ach
Additional error depending on the crest factor (For a non sine-wave input voltage)
Additional error to the ACV(AC+DC)-Ach measurement
Maximum allowable applied voltage
Input
terminal
Range
Maximum display
Resolution
Input impedance
FAST/MED
SLOW1, 2
FAST/MED
SLOW1, 2
V
Hz-
COM A
200 mV
199.9
199.99
100
V
10
V
1 M
12%, 140 pF or less
2000 mV
1999
1999.9
1000
V
100
V
20 V
19.99
19.999
10 mV
1000
V
200 V
199.9
199.99
100 mV
10 mV
700 V
749
749.9
1000 mV
100 mV
Range
Measurement accuracy *1
(% of r digits)
20 Hz -
45 Hz
45 Hz -
100 Hz
100 Hz -
20 kHz
20 kHz -
50 kHz
50 kHz -
100 kHz
200 mV
0.38+14
0.11+14
0.1+14
0.25+15
0.7+24
2000 mV
0.38+14
0.11+14
0.1+14
0.2+15
0.6+24
20 V
0.38+14
0.11+14
0.1+14
0.2+15
0.6+24
200 V
0.38+14
0.11+14
0.1+14
0.2+15
0.6+24
700 V
0.38+10
0.11+10
0.1+10
–
–
*1:
For sine-wave input When RATE is set to FAST or MED, the digit term is reduced
to 1/10 of its specification.
1 - 2
0 + 0.05
(% of r % of range)
2 - 3
0 + 0.05
FAST
MED
(% of reading)
20 - 45 Hz
1.24%
0.72%
45 - 100 Hz
0.06%
–
Between V
Hz and COM A terminals
700 Vrms, 1000 Vpeak, 2.2
10
7
V Hz
Between COM A terminal and the chassis
500 V
Between COM A and COM B terminals
200 V