7352 Series Digital Multimeter Operation Manual
6.6.3 ADC Command Reference
6-24
Calculation
dB/dBm
calculation
DB0
dB calculation OFF
DB1
dB calculation ON
For the voltage and current measurement functions, ON
can be set.
DB2
dBm calculation ON
For the voltage measurement function, ON can be set.
DB?
Reply: One of DB0 to DB2
–
KDn
Constant D
n:0.00001E-9 to 999999.E+6
(1)
KDM
Sets the measurement value to constant D.
KD?
Reply: KD
d.dddddE
dd *3
–
MAX/MIN
calculation
MN0
MAX/MIN calculation OFF
MIN0
MIN1
MIN2
MN1
MAX/MIN calculation ON
MN?
Reply: MN0 or MN1
–
MAX? Reads the MAX value.
*3, *4
Reply: M
d.dddddE
dd
–
MIN?
Reads the MIN value.
*3, *4
Reply: I
d.dddddE
dd
–
AVE?
Reads the AVE value.
*3, *4
Reply: A
d.dddddE
dd
–
AVN? Reads the measurement count.
*3
Reply: AVN
d.dddddE
dd
–
Comparator
operation
CO0
OFF
CO1
ON (If CO1 is set, the single measurement is set.)
CO?
Reply: CO0 or CO1
–
HIn
Constant HIGH
Constant LOW
n:-999999.E+6 to +999999.E+6
Setting resolution:0.00001E-9
(0)
LOn
(0)
HIM
Sets the measurement value to constant HIGH.
LOM
Sets the measurement value to constant LOW.
HI?
Reply: HI
d.dddddE
dd
*3
–
LO?
Reply: LO
d.dddddE
dd
*3
–
*3: The position of the decimal point in the reply is fixed.
*4:
: Header of the Talker format for the measurement data
Item
Command
Description
Valid
command
for the 2nd
display
setting
Initial
value
R6452