7352 Series Digital Multimeter Operation Manual
5.9.5 Comparator Calculation
5-34
The PASS conditions are set in
.
1. Press
in the current set value display state (PASS) to enable the PASS setting
mode.
2. Select the PASS condition by using
and
.
6. Canceling the comparator calculation
The comparator calculation is canceled when any of the following conditions occur.
• The comparator calculation is set to OFF while the comparator calculation is performed.
• The measurement function is changed.
• The *RST command is executed.
7. Changing the measurement range and comparator operation
The comparator function operates continuously even if the measurement range is changed.
The reference values for the judgment are not changed because the values and the unit are set.
Table 5-6 PASS Condition Setting and PASS/FAIL Judgment
Judgment result
Hi
PASS
PASS
FAIL
FAIL
PASS
FAIL
PASS
FAIL
Go
PASS
FAIL
PASS
FAIL
FAIL
PASS
PASS
FAIL
Lo
PASS
FAIL
FAIL
PASS
PASS
PASS
FAIL
FAIL
PASS condition
setting
Panel setting
LGH
– – H
– G –
L – –
L – H
LG –
– GH
– – –
Remote
command
LOP1
LOP0
LOP0
LOP1
LOP1
LOP1
LOP0
LOP0
MIP1
MIP0
MIP1
MIP0
MIP0
MIP1
MIP1
MIP0
HIP1
HIP1
HIP0
HIP0
HIP1
HIP0
HIP1
HIP0
MENU
MENU
CALC
PASS
LOGOHI/_ _ _ _HI/_ _GO_ _/LO_ _ _ _/
LO_ _HI/LOGO_ _/_ _GOHI/_ _ _ _ _ _