7352 Series Digital Multimeter Operation Manual
6.6.3 ADC Command Reference
6-23
Item
Command
Description
Valid
command
for the 2nd
display
setting
Initial
value
R6452
Calculation
NULL
calculation
NL0
OFF
NL1
ON
NL?
Reply: NL0 or NL1
–
KNLn NULL constant setting
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.00001E-9
Note:
The setting is disabled when the NULL calcula-
tion is set to OFF.
(0)
KNL? Reply: KNL
d.dddddE
dd *3
–
Smoothing
calculation
SM0
OFF
SM1
ON
SM?
Reply: SM0 or SM1
–
TIn
Smoothing count
n: 2 to 100 (counts)
(10)
TI?
Reply: TIddd
–
Scaling
calculation
SC0
OFF
SC1
ON
SC?
Reply: SC0 or SC1
–
KAn
Constant A (Zero cannot be set).
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.00001E-9
(1)
KBn
Constant B
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.00001E-9
(0)
KCn
Constant C
n:-999999.E+6 to +999999.E+6
Setting resolution: 0.00001E-9
(1)
KAM
Sets Constant A to the measurement value.
KBM
Sets Constant B to the measurement value.
KCM
Sets Constant C to the measurement value.
KA?
Reply: KA
d.dddddE
dd *3
–
KB?
Reply: KB
d.dddddE
dd
*3
–
KC?
Reply: KC
d.dddddE
dd
*3
–
*3: The position of the decimal point in the reply is fixed.