7352 Series Digital Multimeter Operation Manual
6.6.4 SCPI Command Reference
6-32
NOTE:
When using the SCPI commands via USB, note the following point.
After sending a setting command, send a query command and receive its response.
Sending more than one setting command in succession may cause unintended behavior or an error to
occur.
As query command for setting commands, error check by ":SYSTem:ERRor?" and completion check by
"*OPC?" are generally used.
1. Measurement function and trigger system relevant command
Item
SCPI command
Description
Initial
value
Measurement Measurement
function
[:SENSe]:FUNCtion "<function>"
[,(@<ch>)]
"<function>":
"VOLTage:DC"
DC voltage measurement (DCV)
"VOLTage:AC"
AC voltage measurement (ACV)
"RESistance"
Resistance measurement (2W
)
"CURRent:DC"
DC current measurement (DCI)
"CURRent:AC"
AC current measurement (ACI)
"VOLTage:ACDC"
AC voltage (AC+DC coupling)
measurement (ACV(AC+DC))
"CURRent:ACDC"
AC current (AC+DC coupling)
measurement (ACI(AC+DC))
"DIODe"
Diode measurement (DIODE)
"RESistance:LPOWer" Low power resistance measure-
ment (LP-2W
"CONTinuity"
Continuity test (CONT)
"FREQuency"
Frequency measurement (FREQ)
"VOLTage:BDC"
Bch DCV measurement
(DCV-Bch)
"CURRent:BDC"
Bch DCI measurement (DCI-Bch)
"CURRent:BAC"
Bch ACI measurement (ACI-Bch)
"CURRent:BACDc"
Bch ACI(AC+DC) measurement
(ACI(AC+DC)-Bch)
"TEMPerature"
Temperature measurement (TEMP)
<ch>
1: 1st display
2: 2nd display
[:SENSe]:FUNCtion? [(@<ch>)]
Reply: "VOLT:DC", "VOLT:AC", "RES",
"CURR:DC",CURR:AC", "VOLT:ACDC",
"CURR:ACDC", "DIOD","RES:LPOW",
"CONT", "FREQ", "VOLT:BDC", "CURR:BAC",