User     

     

 

Manual                             

 

 

APPA 607

 

 

Summary of Contents for 607

Page 1: ...User Manual APPA 607 ...

Page 2: ...ys use proper terminals switch position and range for measurements Verify the Meter s operation by measuring a known voltage If in doubt have the Meter serviced Do not apply more than the rated voltage as marked on Meter between terminals or between any terminal and earth ground Only replace the blown fuse with the proper rating as speci fied in this manual Use caution with voltages above 30 Vac r...

Page 3: ...e meter in Ω C mA INSULATION function to measure the voltage of a power supply circuit in equipment that could result in damage the meter and the equipment under test Symbols as marked on the Meter and Instruction manual Unsafe Voltage To alert you to the presence of a potentially hazardous voltage when the Tester detects a voltage 30 V or a voltage overload OL in V mV insulation function The symb...

Page 4: ... Rotary switch for turn the Power On Off and select the function 4 Input Terminal for Insulation function 5 Input Terminal for V Ω A Hz C functions 6 Common Ground reference Input Terminal for all functions expect Insulation function 7 Input Terminal for mA or Common Input Terminal for Insulation function ...

Page 5: ...s When connecting the test leads to the DUT Device Under Test connect the common mA test lead before connecting the live lead when removing the test leads removing the test live lead before removing the common test lead The figures on the following pages show how to make basic measurements Measuring AC DC Voltage ...

Page 6: ...5 Measuring Resistance Continuity Diode Measuring Resistance Continuity Diode Press the Blue button to select the measuring function Testing Diode ...

Page 7: ...6 Testing Continuity The buzzer allows you to quick continuity tests without having to watch the display The buzzer sounds when a short 30Ω is defected Measuring Capacitance ...

Page 8: ...7 Measuring Temperature C F Measuring AC DC Current ...

Page 9: ...oltage 50V 100V 250V 500V 1000V Lock test voltage Press the Hold button to lock the test voltage Press the button again to cancel the lock mode Make the reading stably Press the Blue button to make the reading stably the Smooth appears on the display Press the Blue button again to cancel this mode ...

Page 10: ...ange button to select the test voltage 4 Connect the probes to the circuit 5 Push and hold the Test button to start the test The Test and appear on the display The secondary display shows the test voltage applied to the circuit under test The primary display shows the resistance 6 Keep the probes on the test points and release the Test button The resistance reading appears on the primary display u...

Page 11: ...ary switch to V mV or mA position 2 Press the Hz button to measure the frequency 3 Press the Hz button or change the rotary switch position to exit this function Display Hold In the Hold mode the meter holds the reading To use the Hold mode follow the steps below 1 Press the Hold button to activate Hold mode The Hold appears on the display 2 Press the Hold button Blue button Range button or change...

Page 12: ... also calculate an average of maximum and minimum To use the MIN MAX AVG mode follow the steps below 1 Press the MIN MAX button to activate MIN MAX AVG mode The MIN MAX blinks on the display 2 Press the MIN MAX button to step through the present readings maximum minimum and average AVG 3 Press the MIN MAX button for 2 seconds Blue button Range button or change the rotary switch position to exit Di...

Page 13: ...subtract it To use the Relative mode follow the steps below 1 Turn the rotary switch to Ω or position 2 Press the Hz button to activate Relative mode The 0 blinks on the display 3 Press the Hz button Blue button Range button or change the rotary switch position to exit this function Relative mode for Ω ...

Page 14: ...lect the AC DC mode yourself When you turn the rotary switch to V mV or mA position it defaults to Auto Test mode and the AutoTest appears on the display 1 To enter the Manual Test mode press the Blue button 2 In the Manual Test mode press the Blue button to change the AC DC mode 3 Press the Blue button for 2 seconds to activate Auto Test mode High Frequency Reject mode When the rotary switch in V...

Page 15: ... range yourself When you turn the meter on it defaults to Auto Range mode and the Range appears on the display 1 To enter the Manual Range mode press the Range button The Range disappears on the display 2 In the Manual Range mode press the Range button to increment the range After the highest range the meter returns to the lowest range 3 Press the Range button for 2 seconds to activate Auto Range ...

Page 16: ...amounts To recall the reading on the display press the Store button for 2 seconds to activate Recall mode The data amount shows on the secondary display In the Recall mode you can make the following operation Press the button or button to select the data amount Press the button or button for 2 seconds to search data quickly Press the Blue button to clear all stored data in this function Press the ...

Page 17: ...ironment BUZZER The meter beeps once for every valid key press and beeps twice for every invalid key press Power On Options Press button while turning the meter on from OFF position Blue button Disable APO MAX MIN button Disable auto backlight Store button Clear all stored data Test button Display LCD test frame Range button Default C F reading ...

Page 18: ...ce fuse and the batteries Caution Use only a fuse with the amperage interrupt voltage and speed rating specified Fuse rating 440mA 1000V fast blow fuse Replace the batteries as soon as the low batteries indicator appears to avoid false reading 1 5V x 4 alkaline batteries ...

Page 19: ...low 1 Insert a test probe in the V Ω Hz C input terminal 2 Turn the rotary switch to Ω position and press the Blue button to diode function 3 Insert the probe in the mA input terminal If the display reading is 0L the fuse is bad and should be replaced ...

Page 20: ...ting Ambient Non condensing 10 C 11 C 30 C 80 RH 30 C 40 C 75 RH 40 C 50 C 45 RH Storage Temperature 20 C to 60 C 0 to 80 R H batteries not fitted Temperature Coefficient Add 0 15 x Spec Accy C 18 C or 28 C Measure Samples 3 times per second normal Altitude 6561 7 ft 2000m Safety Complies with EN61010 1 UL61010 1 IEC 61010 1 CAT IV 600V CAT Ⅲ 1000V CAT Application field Compliance to EN 61557 IEC6...

Page 21: ...kHz ACmV 100 00mV 0 9 3d 0 9 3d 1000 0mV 0 9 3d 0 9 3d ACV 10 000V 0 9 3d 1 9 3d 100 00V 0 9 3d 1 9 3d 1000 0V 0 9 3d 1 9 3d 1 HFR ACV 10 000V 0 9 3d 2 9 3d 2 100 00V 0 9 3d 2 9 3d 2 1000 0V 0 9 3d 2 9 3d 2 1 60Hz to 1kHz 2 60Hz to 500Hz DC voltage Measurement Function Range Accuracy DCmV 100 00mV 0 08 3d 1000 0mV 0 08 2d DCV 10 000V 0 08 2d 100 00V 0 08 2d 1000 0V 0 08 2d Over voltage protection ...

Page 22: ...Crest Factor of 2 0 to 2 5 add 2 5 to accuracy For Crest Factor of 2 5 to 3 0 add 4 0 to accuracy CF 3 330V 2 500V AC DC Current Measurement Function Range Accuracy DCmA 100 00mA 0 2 2d 400 0mA 0 2 2d ACmA 100 00mA 1 5 2d 1 400 0mA 1 5 2d 1 1 50Hz to 5kHz Overload Protection mA Input Max 440mA AC Conversion Type Conversion type and additional specification are same as DC AC voltage Maximun Input C...

Page 23: ...0 000MΩ 0 5 2d 40 00MΩ 0 5 2d Resistance Open Circuit Voltage Approximate 0 25V Short Circuit Current Approximate 0 25mA Overload Protection 1000V AC rms or DC Continuity and Diode Measurement Function Range Accuracy Continuity 400 0Ω 0 5 2d Diode 2 000V 0 5 2d Continuity Built in buzzer sounds when measured resistance is less than 30Ω and sounds off when measured resistance is more than 100Ω betw...

Page 24: ...32 0 2192 F 1 2 C 0 0 1200 C 328 0 32 0 F 1 1 C 1 36 F 1 18 F Overload Protection 1000V AC rms or DC Function Range Measuring Time Accuracy 10 000nF 0 7sec 1 2 80d 100 00nF 0 7sec 1 2 20d 1000 0nF 0 7sec 1 2 2d 10 000uF 0 7sec 1 2 2d 100 00uF 0 7sec 1 2 2d 1000 0uF 3 75sec 1 2 2d 10 000mF 7 5sec 1 2 20d 40 00mF 7 5sec 1 2 80d Capacitance ...

Page 25: ... 0MΩ 1 5 5d 2000MΩ 1 5 5d 22 0GΩ 10 3d 1000V Test Voltage vs Maximum Resistance Range 50V 55 0MΩ 100V 110 0MΩ 250V 275MΩ 500V 550MΩ 1000V 22 0GΩ Test Voltage vs Minimum Resistance Range with Test Current 1mA 50V 50KΩ 100V 100KΩ 250V 250KΩ 500V 500KΩ 1000V 1MΩ Test Voltage Accuracy 20 0 Short Circuit Test Current 1mA Auto Discharge Function Discharge time 1 sec for C 1μF Maximum Capacitive Load Ope...

Page 26: ...ccident unauthorized repair alteration contamination or abnormal conditions of operation or handling Any implied warranties arising out of the sale of this product including but not limited to implied warranties of merchantability and fitness for a particular purpose are limited to the above The manufacturer shall not be liable for loss of use of the instrument or other incidental or consequential...

Page 27: ...APPA TECHNOLOGY CORP 9F 119 1 Pao Zong Rd Shin Tien Taipai 23145 Taiwan R O C Tel 886 2 29178820 Fax 886 2 29170848 E MAIL info appatech com http www appatech com ...