background image

3 1/2 Digits Unbeatable Best Buy Multimeters with All-Ranges Protection 

OPERATION MANUAL 

1.  SAFETY RULES 

 

 

This meter is designed and tested in accordance with EN publication 61010-1, pollution 

degree II and installation category (overvoltage category) II 600V. 

 

This meter has been tested according to the following EC Directives 

 

89/336/EEC Electromagnetic Compatibility, EN61326

 

 

73/23/EEC Product safety law of Low Voltage Directive, EN61010-1 

 

This meter is designed to be indoor use at temperature 5

°

C to 40

°

C and altitude up to 

2,000m. 

 

To ensure that the meter is used safely, follow all safety and operating instructions in 

this operation manual.    If the meter is not used as described in this operation manual, 

the safety features of this meter might be impaired. 

 

2.  INTERNATIONAL SYMBOLS 

 

Important information 

 

Diode 

 

see manual 

 

Continuity 

 

AC 

 

Ground 

 

DC 

 

Double insulation 

 

3.  SPECIFICATIONS 

3.1 General Specifications 

Display 

:  3 1/2 digit LCD with max. reading of 1999. 

Polarity 

:  Automatic, (-) negative polarity indication. 

Zero adjustment 

:  Automatic. 

Over range indication 

:  Only the MSD “1” is displayed. 

Power 

:  1.5V AAA battery x 3 

Dimension 

:  85 (W) x 155 (H) x 40 (D) mm. 

Net Weight 

:  Approx. 250g. (Including battery). 

 

3.2 Electrical Specifications 

Accuracies are 

±

 (% of r number of least significant digits) at 23

°

±

 5

°

C, <75% RH. 

DC Voltage 

Range 

Resolution 

Accuracy 

Input 

Impedance 

Overload 

Protection 

200mV 

0.1mV 

±

(1.0%+3) 

10M

 

 

600V DC/ACrms   

 

2V 

1mV 

20V 

0.01V 

200V 

0.1V 

600V 

1V 

AC Voltage 

Range 

Resolution 

Accuracy 

Frequency 

(Hz) 

Overload 

Protection 

200V 

0.1V 

±

(1.2%+3d) 

40~400 

600V DC/ACrms 

600V 

1V 

DC Current 

Range 

Resolution 

Accuracy 

Overload 

Protection 

20mA 

0.01mA 

±

(1.2%+3d) 

250V DC/ACrms 

200mA 

0.1mA 

10A 

(Optional)

0.1A 

±

(5.0%+3d) 

Resistance 

Range 

Resolution 

Accuracy 

Open circuit 

voltage 

Overload 

Protection 

200

 

0.1

 

±

(1.2%+3d) 

<1.0V 

250V DC/AC rms 

<30 sec. 

2k

 

0.001k

 

20k

 

0.01k

 

200k

 

0.1k

 

2M

 

0.001M

 

Diode Test 

 

 

Test Voltage 

Test Current 

Overload Protection 

<3V 

Approx. 1.6mA 

250V DC/AC rms 

<30 sec. 

Continuity 

Test 

 

 

Test range 

Open circuit 

Voltage 

Overload 

protection 

Buzzer sounds when 

resistance value 

<30

 

<3V 

250V DC/AC rms 

<30 sec. 

Phase Sequence 

Indication   

Range 

Phase Sequence Indication 

380ACV 

±

10% 

Correct Phase Sequence 

– Phase Sequence Light “ON” 

Live Wire 

Verification Voltage 

Tested Voltage Range 

100~250V 

Square Waveform 

Output 

Output 

Voltage 

Frequency  Wave Form 

Overload Protection 

Approx. 3V 

50Hz 

Square 

Wave 

250V DC/AC rms 

<30 sec. 

 

 

 

 

 

 

 

4.  PANEL DESCRIPTIONS 

 

 

1. 

Live wire verification indication light   

2. 

LCD display 

3. 

Battery test key 

4. 

Multifunction selector 

5. 

HOLD key 

6. 

Phase verification indication light 

7. 

Input terminal 

 

 

 

5.  OPERATION 
 

!

WARNING 

1)  When measuring voltage ensure that the 

instrument is not connected or switched to a current or resistance, or battery test, or 

diode/ continuity check range. Always ensure that the correct terminals are used for the 

type of measurement to be made. 

2)  Use extreme care when measuring voltage above 50V, especially from sources where 

high energy exists. 

3)  Avoid making connections to “live” circuits whenever possible. 

4)  When performing current measurements ensure that the circuit is not “live” before opening 

it in order to connect the test leads. 

5)  Before performing resistance measurements or diode test, ensure that the circuit under 

test is de-energised. 

6)  Always ensure that the correct function and range is selected. If in doubt about the correct 

range, start with the highest and work downwards. 

7)  Extreme care should be taken when using the instrument to conjunction with a current 

transformer connected to the terminals. High voltage may be produced at the terminals if 

an open circuit occurs. 

8)  Ensure that the test leads and prods are in good condition with no damage to the 

insulation. 

9)  Take care not to exceed the overload limits as given in the specifications. 

 

5.1  DC and AC voltage measurement 

1.  Connect the black test lead to the “COM” terminal and red test lead to the “V” terminal. 

2.  Set the multifunction selector to desired DC V or AC V position and connect the test leads 

across the source or load under measurement. 

 

5.2  DC current measurement 

1.  Connect the black test lead to the “COM” terminal and red test lead to the “mA” terminal 

for measurement up to 200mA. 

2.  Set the multifunction selector to desired current range position. 

3.  Connect the test leads in series with the current source to be measured. 

4.  When the tested current is above 100mA, the input voltage drop might above 1V; hence, 

the test duration cannot exceed 15 seconds!   

!

CAUTION: Max. input overload: 250V rms < 20sec. 

 

5.3  Resistance measurement 

1. 

Connect the black test lead to the “COM” terminal and red test lead to the “Ω” terminal.

 

2.  Set the multifunction selector to desired resistance 

(Ω) range posi

tion. 

3.  Connect the test leads across the circuit to be tested. 

!

CAUTION:    Ensure that the circuit to be tested is “dead”. 

           

Max. input overload : 250V rms < 20sec. 

 

5.4  Diode test 

1.  Connect the black test lead to the “COM” terminal and red test lead to the “

” terminal. 

2.  Set the multifunction selector to 

  (same as    ) position. 

3.  Connect the black and red test leads to the cathode (-) and anode (+) ends of the diode 

to be tested respectively. 

4.  Read the forward voltage drop (junction) value from the display. When the forward biased 

is open, the display will shows overload ‘1’ . 

!

CAUTION: Max. input overload: 250V rms < 20sec. 

 

5.5  Phase Sequence Verification 

1.    Connect the red test lead to the “c” terminal, black test lead to “b” terminal and yellow test 

lead to “a” terminal 

2.  Set the multifunction selector to “600V~/abc” range. 

3.  Connect yellow/black/red test lead to three-phase contact point, If the phase  sequence 

LED lights on, indicating the connection is in correct phase sequence. 

4.  If the phase sequence LED lights off, please swap black (b) and red (c) test leads, if the 

LED remains off which means that there is a missing phase or wrong connection, please 

use Live Wire Verification function to verify each phase again. 

   

 

5.6  Live Wire Verification   
1.  Connect the red test lead to the  “ ”  terminal, 

black test lead to “COM” terminal and hold the black 

test lead (please note: 

do not touch the test pin 

and keep your hands away from the finger guard 

of the test lead!

2.  Set the multifunction selector to  “600V~/abc” 

range. 

3.  Use the red test lead to contact the live wire  to 

be tested.  When there is a live wire, the Live Wire 

Verification lights will turn on. 

 

!

Reviews: