6243/44 DC Voltage Current Source/Monitor Operation Manual
A.2 Error Message List
A-4
Self-test
+350
IM 32
A ZERO
261 IM 32
A ZERO test error.
+351
IM 320
A ZERO
IM 32
A ZERO test error.
+352
IM 3.2 mA ZERO
IM 3.2 mA ZERO test error.
+353
IM 32 mA ZERO
IM 32 mA ZERO test error.
+354
IM 320 mA ZERO
IM 320 mA ZERO test error.
+355
IM 2 A ZERO
IM 2 A ZERO test error (6243).
+355
IM 3.2 A ZERO
IM 3.2 mA ZERO test error (6244)
+356
IM CMV
IM CMV test error.
+357
OVL Check
OVL detection check error
+358
IM 10 A ZERO
IM 10 A ZERO test error
Hardware failure
+360
Fan Stop
Cooling fan failure
+361
Over Heat
Internal circuit overheat
+362
Source Unit
The source unit malfunctions
Execution error
+201
Calibration
Calibration error
+210
Sweep Data>5000 (Linear)
The number of steps has exceeded 5000.
+211
|Start|>|Stop| (Log)
In a log sweep, the starting value is greater than the stop value.
+212
Start, Stop Polarity (Log)
In a log sweep, the starting and stopping polarities are different.
+213
Power Over
Source limiter setting exceeds the output range.
+214
Start, Stop =0 (Log)
In a log sweep, the staring or stopping value is 0.
+215
Step =0 (Linear)
The step value for a linear sweep is 0.
+216
Start
0 at FixRange (Log)
For a log sweep, the fix range starting value is 0.
Command error
+101
Command Execution
GPIB command execution error.
GPIB argument range error.
+102
Command Syntax
GPIB command syntax error
An unrecognized command is received.
Source / measurement
message
None
OSC/osc
Oscillation measurement / oscillations detected.
None
RP/
rp
Reversed polarity source measurement / reversed polarity
source connections detected
None
LM/
lm
Limit measurement / Limiter detected
None
Over Range
The measurement value has exceeded the range.
None
Over Load
Over load
None
HI
Comparative calculation result HI
None
GO
Comparative calculation result GO
None
LO
Comparative calculation result LO
External single line
signal test
None
Failed 01,03,05,07,09,11,13,15
SYNC OUT or TRIGGER IN malfunction
None
Failed 02,03,06,07,10,11,14,15
OPERATE OUT or BUSY IN malfunction
None
Failed 04 to 07, 12 to 15
BUSY OUT or OPERATE IN malfunction
None
Failed 08 to 15
COMPLETE OUT or OPERATE IN malfunction
Table A-2 Error Message List (2 of 2)
Category
Error
No.
Display message
Description