6243/44 DC Voltage Current Source/Monitor Operation Manual
2.3.2 Battery Charge Test and Discharge Test
2-38
Connecting the DUT
8. Connect the leads using 4-wire as shown in Figure 2-12 to eliminate voltage drop
caused by the leads.
9. Connect the
COMPLETE OUT
terminal and the
OPERATE IN
terminal on the
rear panel using the BNC-BNC cable A01036.
The charging and discharging are completed automatically.
Figure 2-12 Connections for Battery Charge and Discharge Test
Charge Test
10. Press the
MENU
key. Select
TIME
in the parameter group by rotating the Data
knob.
Press the key and select the
Period
parameter by rotating the Data knob.
Press the key and press the
1, 0, 0, 0,
and
ENTER
keys.
Period is set to 1000.0 ms.
Press the key twice or press the
MENU
key, and select
SOURCE
in the
parameter group by rotating the Data knob.
Press the key and select the
Remote Sensing
parameter by rotating the Data
knob.
Press the key and select
4 Wire
by rotating the Data knob.
Remote Sensing is set to 4 Wire.
Press the key twice or press the
MENU
key, and select
COMPARATOR
in the
parameter group by rotating the Data knob.
Press the key and select the
Compare
parameter by rotating the Data knob.
Press the key and select
ON
by rotating the Data knob.
Compare is set to ON.
Press the key and select the
Upper Data
parameter by rotating the Data knob.
Press the key and press the
2, 0, 0, 0, mA,
and
ENTER
keys.
Upper Data is set to +2000.00 mA.
Press the key and select the
Lower Data
parameter by rotating the Data knob.
Press the key and press the
1, 0, 0, mA,
and
ENTER
keys.
Lower Data is set to +100.000 mA.
Press the key twice or press the
MENU
key, and select
EXT.SIGNAL
in the
parameter group by rotating the Data knob.
+
-
Rear Panel
Front Panel