6243/44 DC Voltage Current Source/Monitor Operation Manual
4.6 Measurement Function
4-29
4.6
Measurement Function
4.6.1
Measurement Function
Regardless of the source function, the measurement function can be selected as voltage measurement or
current measurement.
4.6.2
Measurement Ranging
The measurement ranging types and their functions are shown below:
NOTE:
If the source value or limiter value is set to the full-scale value, errors may cause the measurement value
to exceed the full-scale value resulting in Over Range display.
Range
Function
Range before
measurement
Operation during
measurement
Fixed range
Measure in the Source range or
Limiter range
Limiter range or Source
range
Fixed range
Auto range
A range to maximize the effec-
tive digits for measured value is
selected
Limiter range or source
range
Auto ranging *
*
For voltage source current measurement (VSIM) and current source voltage measurement
(ISVM), the measurement range becomes the auto range in which the limiter range is the maxi-
mum range.
For voltage source voltage measurement (VSVM) and current source current measurement
(ISIM), the measurement range is identical with the source range and the auto range does not
function.