4.3 Operation in the Pulse Source Mode
4-12
4.3
Operation in the Pulse Source Mode
Table 4-3 shows the operation in the Pulse source mode.
Table 4-3 Operation in the Pulse Source Mode
Condition
change
Sampling
mode
Operation
Description
Remarks
Changing the
source value in
the range
mid-operation
Free-run
(a)
Continuous
pulse
generation in
specified
Tp + Tds
Tp: Period
Tw: Pulse width
Tm: Measurement
time
Integration time
+ Measurement data
Processing time
Tds: Source delay
time
Tcn : Operate process-
ing time
Trc : Processing time
for the range
change
Fore restrictions of Tp,
Tds, Tw, Tm and
Td, refer to
4.9.1.
Hold
(b)
Single pulse
generation
after elapsing
Tds from the
trigger input
Changing from
Standby to
Operate or
changing the
range
Free-run
(c)
Pulse genera-
tion after
elapsing Tcn,
Trc, and Tds
from the time
when specify-
ing Operate or
changing the
source value.
Hold
(d)
Single pulse
generation
after elapsing
Tds from the
trigger input
Tp
Tw
Td
Tm
Previous
value
Base
value
Tds
Tp
Tw
Td
Tm
Tds
Tp
Tw
Td
Tm
Source value change
Tds
Tp
Tw
Td
Tm
Previous
value
Source value
change
Trigger
Standby status
or base value
Operate
Source value
change
Tcn, Trc
Tds
Td
Tm
Tw
Tp
Tds
Tp
Tw
Td
Tm
Standby
status or base
value
Operate
Source value
change
Trigger
Base
Tcn, Trc
Tds
Tp
Tw
Td
Tm