CZ 153
0.001 g
150 g
0.001 g
±0.003 g
2 mg
100 g
(F1)
Ø 80
CZ 203
0.001 g
200 g
0.001 g
±0.003 g
2 mg
200 g
(F1)
Ø 120
CZ 302
0.01 g
300 g
0.01 g
±0.03 g
20 mg
200 g
(F1)
Ø 120
CZ 602
0.01 g
600 g
0.01 g
±0.03 g
20 mg
500 g
(F1)
Ø 120
CZ 1202
0.01 g
1200 g
0.01 g
±0.03 g
20 mg
1000 g
(F1)
Ø 120
CZ 2002
0.01 g
2000 g
0.01 g
±0.03 g
20 mg
2000 g
(F1)
Ø 120
10, 20, 50, 100, 200
g, ct, lb, oz, d, ozt, dwt, mo, tl h, tl c, tl t, t, bt, n
2 hours
3 sec.
+ 5° C .... + 35° C
max. 80 % (not condensing)
2.6 kg
180 x 220 x 85mm
9 V / 800 mA
Aczet
Readability (d)
Weighing range (max)
Reproducibility
Linearity
Smallest part weight for
parts counting
Reference quantities at
parts counting
Weighing Units
Recommended adjustment
weight, not added (class)
Warm-up time
Stabilization time (typical)
Operating temperature
Humidity of air
Weighing plate, stainless
steel (mm)
Gross Weight
Overall Dimension (WxDxH)
Voltage
1
1. Technical Data
9.4 Continuous
data
output
Head line 1 : ST =stable,
US=instable
Head line 2: NT=net,
GS=gross
9.5
Remote control instructions
The remote control commands are sent from the remote control unit to the balance as ASCII code.
After the balance having received the s/w/t commands, it will send the following data.
T
Function :
Tare balance
Z
Function :
Set balance to zero
W
Function :
Weighing value for the weight (stable or unstable) is sent via the
RS 232 interface
S
Function :
Stable weighing value for the weight is sent via the RS 232 interface
P
Function :
In counting mode the number of the pieces to be counted is displayed
in “Pcs”
38
-Header 1-
-Header 2-
--Weighing Data--
-/U
k
g
CR
LF
,
Weighing Unit
Final character