13
15
5.8.2
Linearization of non NTEP models (low resolution) and NTEP models
• Press , “Pin” appears
• Wait for the stability display and the signal sound
• “LoAd 2” appears
• Remove weight 1 and put on the weight 2
• After the stability display and the signal sound „LoAd 3“ appears;
remove weight 2 and put on weight 3
• After the stability display and the signal sound the balance carries
out a selftest. Remove the weight during selftest. The balance
changes into weighing mode. Now the linearization is concluded
successfully.
• Press repeatedly until “tECH“ appears
In case of verifiable models press the adjustment switch on the lower side of the balance!
• Start balance by pressing . While the balance carries out a
self test, press until “F1 Unt” appears
MODE
MODE
Operation
Balance Display
• Press a new , “Pin” appears again
• Press , and successively, “P1 Lin” appears
Tare
• Press successively , and , “LoAd 0” appears;
after the apparition of the stability display and the signal sound,
“LoAd 1” will be displayed
• Put on weight 1 (see table chap.6.8.1)
7.3.2 NTEP
models
• Remove the load
• Press , the balance goes to weighing mode.
MODE
MODE
By pressing the key several times, the reference quantities 10,
20, 50, 100, 200 can be set
Operation
Balance Display
• Press
Appears the reference quantity “10”.
• Start balance by pressing
The balance will carry out a self-test.
Wait for „0.0“ display.
• Place as many parts to count on the weighing plate as the set
reference quantity requires.
• Confirm by pressing
The balance is now in parts counting mode and counts all units on
the weighing plate.
• Press
The balance is in weighing mode and displays the weight of the
counted parts.
24