5.Unpacking, Setup and Commissioning
5.1 Installation
Site,
Location
of
Use
The balances are designed in a way that reliable weighing results are achieved in common
conditions of use.
You will work accurately and fast, if you select the right location for your balance.
Therefore, observe the following for the installation site:
• Place the balance on a firm, level surface;
• Avoid extreme heat as well as temperature fluctuation caused by installing next to a
radiator or in the direct sunlight;
• Protect the balance against direct draughts due to open windows and doors;
• Avoid jarring during weighing;
• Protect the balance against high humidity, vapors and dust;
• Do not expose the device to extreme dampness for longer periods of time. Non- permitted
condensation (condensation of air humidity on the appliance) may occur if a cold appliance is
taken to a considerably warmer environment. In this case, acclimatize the disconnected
appliance for ca. 2 hours at room temperature.
• Avoid static charge of goods to be weighed or weighing container.
Major display deviations (incorrect weighing results) may be experienced should electromagnetic
fields (e.g. due to mobile phones or radio equipment), static electricity accumulations or instable
power supply occur. Change location or remove source of interference.
5.2 Unpacking
Carefully remove the balance from the packaging and setup balance at the intended workstation.
32
8.5 Menu Map (NTEP models)
Operation
Balance Display
Not available
Background lighting on/automatic/off
• EL on: background lighting on
• EL Au: background lighting switches on automatically
• EL off: background lighting off
RS -232 interface
S 232:
• P Prt: connect with printer
• b XXXX: Baud rate can be selected between 9600, 600, 1200,
2400, 4800
• LP-50: not documented
•
EnG: Selection of Language English
•
Chi:
Selection
of
Language
Chinese
• tP: Standard printer setting
• P Cont: Continuous printout
• P AUto: autom. Printout
• P ASk: Data output via remote control commands
S USb: not documented