13. 3 ELECTROMAGNETIC IMMUNITY / RADIOFREQUENCY MOBILE EQUIPMENT
The P
IEZOTOME
2 is intended for use in the electromagnetic environment specified in the table below.
The user and/or installer must ensure that the device is used in such an electromagnetic environment.
Immunity test
IEC
60601
test
level
Compliance
level
Electromagnetic environment - guidance
Portable and mobile radiofrequency communications devices must not be used near the P
IEZOTOME
2
(including its cables) at a distance less than that recommended and calculated according to the
frequency and power of the emitter.
Conducted
disturbance,
radiofrequency fields.
IEC61000-4-6
3 V/m
150 KHz
to 80
MHz
3 V/m
Recommended separation distance:
d = 1.2
P
Radiated
radiofrequency
electromagnetic field.
IEC61000-4-3
3 V/m
80 MHz
to 2.5
GHz
3 V/m
d = 1.2
P
80MHz to 800MHz.
d = 2.3
P
800MHz to 2.5GHz.
Where P is the maximum power rating of the
emitter in watts (W) according to the
manufacturer's specifications and d is the
recommended minimum separation distance in
meters (m).
The electromagnetic field strengths of fixed radiofrequency emitters, as determined by an
electromagnetic environment measurement (a), must be less than the compliance level in each
frequency range (b).
Interference may occur near equipment marked with the symbol below:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: -These specifications may not be applicable in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and persons.
(a): -The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for
mobile telephones (cellular/cordless), mobile radios, amateur radio, AM/FM radio broadcasts and TV
broadcasts cannot be determined exactly by theory. To assess the electromagnetic environment due to
fixed radiofrequency emitters, an electromagnetic environment measurement must be made. If the
measured radiofrequency field strength in the immediate environment where the product is used
exceeds the compliance level specified above, the performance of the product must be tested to verify
whether it conforms to the specification. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the product.
(b): -In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3
V/m.
13. 4 RECOMMENDED SEPARATION DISTANCES
P
IEZOTOME
2 is intended for use in an electromagnetic environment in which radiated radiofrequency
disturbances are controlled.
The P
IEZOTOME
2 user and/or installer can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile radiofrequency communications equipment emitters
and the P
IEZOTOME
2, according to the maximum output power of the equipment, as recommended in the
table below.