To avoid this happening, place the items in individual sterilizable bags or in a sterilization box.
To maintain the sterility or asepsis of the accessories (cords, handpieces, etc.), store them in sealed
bags or containers suitable for medical practice.
The maintenance and/or sterilization instructions below must be applied before each use of the
device.
Avoid using cleaning and disinfection agents containing flammable substances.
If this is not possible, make sure that all the agent has evaporated and that there is no combustible
matter on the device or its accessories before it is switched on.
Sterilized items must be allowed to dry and cool to room temperature before they are reused.
Before sterilization, check that your autoclave is clean and that the water used is of adequate
quality.
After each sterilization cycle, remove the items from the autoclave immediately in order to
minimize the risk of metal corrosion.
11. 1 IRRIGATION LINE MAINTENANCE
Sterile irrigation line
The irrigation lines supplied by SATELEC are disposable and must always be discarded after use. Re-use
of an irrigation line may result in contamination of your patients and expose you to de facto liability. Do
not attempt to resterilize the single-use irrigation lines. Do not attempt to modify the irrigation lines.
Sterilizable irrigation line
Refer to the user's manual delivered with the equipment.
11. 2 DEVICE MAINTENANCE
Important:
Never use an abrasive cleaning agent on the device.
Do not use sprays or liquids to clean and disinfect the P
IEZOTOME
2 control unit.
The device, the brackets and the multifunction footswitch are not sterilizable.
The P
IEZOTOME
2 control unit, the control footswitch and the bracket must always be cleaned and
disinfected after each procedure, using alcohol, disinfectant agents or disinfectant wipes for dental
practices, such as S
EPTOL
™ wipes.
Regular monitoring of the P
IEZOTOME
2 unit is necessary in order to detect any problem.
It is important to keep the control unit ventilation vents clean in order to avoid abnormal heating.
11. 3 SCALER CORD MAINTENANCE
Important:
Never use an abrasive cleaning agent on the cords.
The cords must not be immersed or disinfected using agents containing acetone, chlorine or bleach.
The cords must not be cleaned in an ultrasonic bath.