Weight: approximately 3.5 kg
LCD screen
:
Height: 86 mm
Width: 115 mm
Scaler handpiece cord
: 2 000 mm (2 900 mm optional)
P
IEZOTOME
handpiece cord
: 2 000 mm (2 900 mm optional)
Temperatures
Operation: +10°C to +40°C
Storage: -20°C to +70°C
Humidity
Operation: 30% to 75%
Storage: 10% to 100% including condensation
Atmospheric pressure
Between 500 hPa and 1060 hPa
IV - -INSTALLATION / FIRST USE
4. 1 UNPACKING THE DEVICE
Upon receipt of the device, check for any damage caused in transit.
Contact your supplier if necessary.
4. 2 RECOMMENDATIONS
Have your P
IEZOTOME
2 connected to the power electricity supply by a certified dental installation
technician.
The electrical connection of the P
IEZOTOME
2 must comply with the applicable standards in your country.
Warning:
To avoid any risk of electric shock, this device must be connected to an electrical power supply
with a protective ground.
4. 3 INSTALLATION
Important:
Do not place the P
IEZOTOME
2 close to or on top of another device.
Do not place the power cord and the footswitch cord in a wire cover or in a cable gland.
- Place the control unit in position on a fixed surface, horizontal or with a slope of no more than 5°.
- Check that the power switch (Fig. 2-5) is in position O (OFF).
- Connect the power cord to the device.
- Connect the power cord to a power socket with a ground pin.
- Connect the control footswitch to the footswitch connector (Fig. 2-2).
- Position the footswitch so that it is easily accessible for your feet.