If necessary, modify the program power level (1 to 5) using buttons
(Fig. 3-6) and (Fig. 3-7).
The power level of the program concerned is shown as a number and a bar-graph
(Fig. 4-13).
Modify the footswitch type if necessary by pressing button
(Fig. 3-10) to toggle between ON/OFF
and progressive
.
Important:
In progressive mode, for programs D1 to D3, the delivered power ranges between power level 1 of
program D3 and the configuration defined by the user (program and power level).
In progressive mode, the power delivered by program D4 ranges only between levels 1 and 5,
according to the value defined by the user.
If necessary, select handpiece light function ON/OFF by pressing button
(Fig. 3-11).
Confirm the new settings by pressing button
(Fig. 3-8).
The saving of the various settings is confirmed by an audible signal and a visual signal (button flashing).
Repeat this sequence if necessary in order to adjust all the programs.
Caution: The tips designed for P
IEZOTOME
2 and I
MPLANT
C
ENTER
2 are not compatible with P
IEZOTOME
and I
MPLANT
C
ENTER
, and vice versa.
7. 6 NEWTRON FUNCTION
Important:
A scaler cord and its N
EWTRON
handpiece must be connected to the device in order to be able to
configure the various settings.
The N
EWTRON
function settings can be customized by the user.
Select the N
EWTRON
function by pressing button
(Fig. 3-12) or by pressing the footswitch button
(Fig. 8-5).
On the LCD screen, the N
EWTRON
tab must be highlighted.
Select the program for which settings are to be adjusted by pressing buttons
,
,
or
(Fig. 5-16) or by pressing the footswitch button (Fig. 8-3).
Adjust the power level according to the TipBook recommendations.
If necessary, modify the program power level (1 to 10) using buttons
(Fig. 3-6) and (Fig. 3-7).
The power level of the program concerned is shown as a number and a bar-graph
(Fig. 5-15).
Modify the footswitch type if necessary by pressing button
(Fig. 3-10) to toggle between ON/OFF and
progressive.
If necessary, select handpiece light function ON/OFF by pressing button
(Fig. 3-11).
Confirm the new settings by pressing button
(Fig. 3-8). The saving of the various settings is
confirmed by an audible signal and a visual signal (flashing button).
Repeat this sequence if necessary in order to adjust all the programs.