No ultrasonic function
Incorrect tip tightening.
Retighten the tip with the
wrench.
Faulty connector contact.
Clean the connector contacts.
Handpiece cord wire cut.
Return to SATELEC after-sales
to replace the cord.
No light
Handpiece light ring absent.
Install the light ring.
Faulty light ring.
Replace the light ring.
Faulty light ring connector contacts.
Clean the light ring connector
contacts.
Light ring the wrong way around.
Position the light ring
according to the foolproofing
stud.
Faulty handpiece and/or cord
connector contacts.
Clean the handpiece and/or
cord connector contacts.
Other.
Contact your installer-fitter.
Leak between the N
EWTRON
handpiece and the handpiece
cord
Worn handpiece 1.15x1 seal
Replace the seal (kit F12304).
Liquid leak in the irrigation
pump
Broken tube in the irrigation line
cassette.
Replace the irrigation line.
XIII - ELECTROMAGNETIC COMPATIBILITY
Important:
The power cord, the ultrasonic handpiece cords and the multifunction footswitch cord must be kept
away from each other.
P
IEZOTOME
2 requires special precautions to be taken with regard to electromagnetic compatibility.
It must be installed and prepared for use as described in chapter 4.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere
with the P
IEZOTOME
2.
The recommended separation distances in this paragraph must therefore be complied with.
P
IEZOTOME
2 must not be used near or on top of another device.
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have the
consequence of increasing the emissions or decreasing the immunity of the P
IEZOTOME
2.