a) Cleaning and disinfection
The cords must always be cleaned, disinfected and sterilized after each procedure, using alcohol,
disinfectant agents or disinfectant wipes for dental practices, such as S
EPTOL
™ wipes.
Disconnect the N
EWTRON
handpiece from its cord before cleaning, disinfection and sterilization.
b) Sterilization
The cords can be sterilized in an autoclave under the following conditions:
- Autoclave: Class B
- Sterilization temperature: 134°C at 2 Bar.
- Time at sterilization temperature: 18 minutes.
CAUTION: Not all autoclaves can reach 134°C. Not all autoclaves draw a pre-vacuum. Please refer to
your autoclave manufacturer for specific sterilization instructions.
11. 4 ULTRASONIC HANDPIECE MAINTENANCE
After each use, the handpiece irrigation circuit must be rinsed with distilled or demineralized water for
20 to 30 seconds.
Cleaning and disinfection
:
The handpieces must always be cleaned, disinfected and sterilized after each procedure, using alcohol,
disinfectant agents or disinfectant wipes for dental practices, such as S
EPTOL
™ wipes.
For handpiece sterilization, refer to the manual for each product.
11. 5 TIP MAINTENANCE
a) Lifetime
Tip shape and weight are the determining characteristics for obtaining maximal efficiency of the
ultrasonic generator.
Careful monitoring of these two characteristics by the user will ensure that optimal performance of the
device is maintained.
Consequently, you are strongly advised not to alter the structure of the tips by filing or twisting them or
modifying them in any other way.
Similarly, through normal wear, tip aging leads to alteration of its characteristics. Always replace a tip
that has been damaged by wear or by accidental impact (dropped, bent, etc.).
Use the
T
IPCARD
to check scaler tip wear. Replace routinely-used tips at least once a year.
b) Pre-disinfection / chemical cleaning
To be carried out immediately after treatment, wearing thick gloves. Immerse in an ultrasonic bath
(enzymatic, quaternary ammonium base detergent / disinfectant solution), carefully respecting the
concentration and contact time recommended by the manufacturer of the solution.
Use a solution that bears the EC marking or that is in compliance with any standard required by the
national regulations.
Rinse under running water for at least 30 seconds.