Maintenance and Repair
Maintenance and Repair
13
Maintenance and Repair
For this measuring device, calibration is imperative. It should be included in the
measuring equipment monitoring of the enterprise.
Always keep the A1040 Mira 3D clean and maintained state. Avoid unnecessary
force when using the device and always dry and clean the device before storage.
For storage use the provided box with the foam inlay. Also clean and dry the insides
of the box before the device is stored inside. See chapter Troubleshooting for first
steps if you are in need of help or find any malfunctions.
Preventive Maintenance
Frequency
Parts Concerned
Action
Before each
use
Main
Device:
Housing
Check for mechanical damage.
Transducer
mechanics
Check for smooth
operation,mechanical damage and
cleanliness
Never try try to remove the protection
bellow .
Never try to lubricate the the
suspension system.
Transducer
tips
Check for mechanical damage and
cleanliness
Protection
bellows
Check for mechanical damage and
cleanliness
Information
Sticker
Check for completeness and
readability.