35
A1040 MIRA 3D
© 2021 Acoustic Control Systems – ACS Group
Maintenance and Repair
Maintenance and Repair
USB-C socket
Check for mechanical damage and
cleanliness.
Control Unit
Check for mechanical damage and
cleanliness.
Before
functionality
check
Transducer test plate
Check for mechanical damage and
cleanliness.
After each use
Main
Device:
Housing
Check for mechanical damage.
Transducer
mechanics
Check for smooth
operation,mechanical damage and
cleanliness
Never try try to remove the protection
bellow .
Never try to lubricate the the
suspension system.
Transducer
tips
Check for mechanical damage and
cleanliness / Clean the transducer tips
with with a small, soft brush and a
little moistened cloth.
Protection
bellows
Check for mechanical damage and
cleanliness / clean the Protection
Bellows with a small, soft brush and a
little moistened cloth.
Information
Sticker
Check for completeness and readability
/ Clean with a little moistened cloth /
Ask service for replacement of missing
or unreadable stickers.
Control Unit
Check for mechanical damage and
cleanliness / Clean with a little
moistened cloth
Once per year
Complete A1040 Mira
3D system
Inspection and calibration by ACS or
authorized dealer.
Control procedures and test equipment
Correct functionality is inspected by means of the test plate contained in the set.
The procedure is detailed at
.