Setup
Setup
Figure 24: A1040 MIRA 3D PRO connection
23.2 Automatic transducer testing
The delivery kit includes a
test block
. The block is a plexiglass plate with
a slightly
rough surface. It is used to perform the functionality test of each array transducer.
23.2.1
A1040 MIRA 3D
This section describes the step-by-step procedure for transducer testing of A1040 MIRA
3D.
Step 1
Put the sample on a flat, dry and stable surface.
Step 2
Switch on the tomograph, establish the connection to the hotspot and start the app.
Refer to the details for the
getting started
procedure on
. Open
setup
mode
and configure the instrument as master.
Step 3
Activate the
testing procedure
by pressing the
button. After activation the
app shows the following window. Put the instrument on the sample and press it
firmly.