The device is equipped with a semi-automatic sensor calibration (
SENSu
/
SENSa
). A switching output
operates the trimming resistor, which exists in some sensors. An adjustment of offset and final value takes
place, after which the sensor can be used directly. Depending on parameterisation, the calibration can be
realized via the 4th key or via the digital input. It is possible to key during the calibration steps. So, reference
signals can be connected manually. However the calibration will be interrupted after 30 seconds.
Press actuator for minimum 3
seconds.
For minimum 1 measuring cycle, minimum
1s or maximum per actuator.
Current measuring range value,
e.g.
0.061
For minimum 1 measuring cycle or
with actuator.
For minimum 1 measuring cycle, minimum
1s or maximum per actuator.
Current measuring range value.
Switch-off/switch-on optional switching
output, e.g. 7,132
For minimum 1 measuring cycle,
minimum 1s or maximum per actuator.
Optional: switch-off switching output and display
for minimum 1 measuring cycle or minimum 1s
holding time.
Back to regular operation mode.
For calibration
problems or 30
seconds holding
time.
Back to regular operation mode, after
1s or 1 measuring cycle.
Activate optional switching output.
For calibration
problems or 30
seconds holding
time.