T-APM - Apex Premier F
UME
H
OOD
M
ONITOR
T-APM 0133MAN00015 - Rev J ECN - JUL 2017
Page 30 of 61
Configuration of the APM
Calibration Menu Selection
The
CALIBRATION MENU
can be selected at any time from the
CONFIGURATION MENU
screen, see
[SCREEN
1G]
. Press the
CALIBRATION MENU
button to go to the
CALIBRATION
section of the APM (reference the
CALIBRATION
section of this manual for more information).
Restore Factory Defaults
Another helpful option on the APM is the ability to restore the entire APM to factory defaults at any time.
This is desirable if the APM setup that was performed was incorrect. Sometimes starting over again from
“scratch” is the best way to make sure nothing was missed when re-configuring or re-calibrating a unit.
Note: that this selection will permanently wipe out all field changes that were made to the APM including
all calibrations, set points, passwords, and adjustments. The result is that the APM is returned to the
same state it was in when it was first powered up right out of the box. For a complete list of all default
values, see Appendix I.
To return the APM to factory defaults, select the
RETURN TO FACTORY DEFAULTS
option from the
CONFIGURATION MENU
, see
SCREEN [1H]
. The APM will prompt again for confirmation, see
[SCREEN 2H]
.
Press
ACCEPT
to return the APM to default values, press
CANCEL
to exit out.
Calibrate Touch Screen
The touch screen is factory calibrated and should not require re-calibration. However, if need be, it can
be re-calibrated as follows:
Step 1
: Press and hold all three buttons on the face of the APM until a small beep is heard, this takes
about 5 to 6 seconds.
Step 2
: Next release the center button only. The touch screen calibration will start automatically, just
hold onto the outer two buttons until the screen changes
SCREEN [20K].
Step 3:
Simply follow the directions on the screen and touch the small dot displayed on the upper left
hand side of the screen
SCREEN [21K]
. Next touch the small dot located on the lower right hand side of the
screen
SCREEN [22K]
and that’s it! Use of a stylus is highly recommended for the calibration of the touch
screen.
Exit Configuration and Calibration
At any time, selecting the
EXIT
selection from the
CONFIGURATION MENU
and pressing it will put the APM
back into its normal operational mode.