APPENDIX F:
Configuration and Calibration
Screens
SCREEN 2M
CONTRAST ADJUST
INCREASE
DECREASE
ACCEPT
CANCEL
SCREEN 3N
3 POINT CALIBRATION
SELECT LOW FACE
VELOCITY
+
-
60
ACCEPT
RESET
FPM
3 POINT CALIBRATION
OPEN SASH TO THE FULL
OPEN POSITION, WITH THE
STOPS DEFEATED
PRESS TO ACKNOWLEDGE
SCREEN 1N
3 POINT CALIBRATION
PERFORM TRAVERSE
ACROSS ENTIRE FACE AREA
OF FUME HOOD
PRESS TO ACKNOWLEDGE
SCREEN 2N
3 POINT CALIBRATION
SET SASH TO
MID-OPEN POSITION
PRESS TO ACKNOWLEDGE
SCREEN 4N
SCREEN 6N
3 POINT CALIBRATION
SELECT MID FACE
VELOCITY
+
-
113
ACCEPT
RESET
FPM
3 POINT CALIBRATION
PERFORM TRAVERSE
ACROSS
MID-OPEN POSITION
PRESS TO ACKNOWLEDGE
SCREEN 5N
3 POINT CALIBRATION
SET SASH TO
MINIMUM POSITION
PRESS TO ACKNOWLEDGE
SCREEN 7N
3 POINT CALIBRATION
PERFORM TRAVERSE
ACROSS MINIMUM
POSITION
PRESS TO ACKNOWLEDGE
SCREEN 8N
SCREEN 9N
3 POINT CALIBRATION
SELECT HIGH FACE
VELOCITY
+
-
176
ACCEPT
RESET
FPM
SCREEN 1P
FIELD
CALIBRATION
REQUIRED
SCREEN 2P
WARNING
FACE VELOCITY
BELOW RANGE
SCREEN 3P
WARNING
FACE VELOCITY
ABOVE RANGE