Calibration of the APM
Step 4b: Verify New Reading
The next displayed screen,
[SCREEN 18K]
, will show a face velocity reading for the given position of the
sash at that moment. It is a live reading and varying the sash position will vary the reading as expected.
If the new readings for the high velocities (sash in a minimum position) are correct as displayed, press
ACCEPT
, if not, press
RETRY
to try again or
CANCEL
to quit. Note: the low face velocity readings do not
change as the only adjusted point was for the high velocity readings.
Step 5b: Low Velocity Calibration Point Adjustment
After selecting
LOW VEL. CAL POINT
, the APM will ask for an offset value to move the calibration point
by
[SCREEN 17K]
. This value will be added (or subtracted when a negative number is entered) to the
calibration point obtained when the two point calibration was originally performed. Using the plus and
minus buttons [
or
] enter a positive or negative offset value and press
ACCEPT
.
Step 6b: Verify New Reading
The next displayed screen,
[SCREEN 18K]
, will show a face velocity reading for the given position of the
sash at that moment. It is a live reading and varying the sash position will vary the reading as expected.
If the new readings for the low velocities (sash in a full open position) are correct as displayed, press
ACCEPT
, if not, press
RETRY
to try again or
CANCEL
to quit. Note: the high face velocity readings do
not change as the only adjusted point was for the low velocity readings.
Step 7b: Exiting Out of the Calibration Adjustment Menu
On the
TWO POINT CALIBRATION ADJUSTMENT
main screen,
[SCREEN 15K]
select
EXIT
to go back to the
CALIBRATION MAIN MENU
,
[SCREEN 1K]
.