APPENDIX I:
Default Values of APM
Variable Name
Factory Default
Range for Variable
High Alarm Mode
Occupied
Occupied or Unoccupied
Occupied High Alarm Mode
Disabled
Disabled or Enabled
Occupied High Alarm Set Point
Disabled
100 to 500 FPM
Occupied High Alarm Delay
Disabled
0 to 60 seconds
Occupied High Alarm Caution Band
Disabled
0 to 99 FPM
Unoccupied High Alarm Mode
Disabled
Disabled or Enabled
Unoccupied High Alarm
Disabled
100 to 500 FPM
Unoccupied High Alarm Delay
Disabled
0 to 60 seconds
Unoccupied High Alarm Caution Band
Disabled
0 to 99 FPM
Low Alarm Mode
Occupied
Occupied or Unoccupied
Occupied Low Alarm Mode
Disabled
Disabled or Enabled
Occupied Low Alarm Set Point
Disabled
0 to 99FPM
Occupied Low Alarm Delay
Disabled
0 to 60 seconds
Occupied Low Alarm Caution Band
Disabled
0 to 99 FPM
Unoccupied Low Alarm Mode
Disabled
Disabled or Enabled
Unoccupied Low Alarm
Disabled
0 to 60 seconds
Unoccupied Low Alarm Delay
Disabled
0 to 99 FPM
Unoccupied Low Alarm Caution Band
Disabled
Disabled or Enabled
Purge Button
Disabled
Disabled or Enabled
Display Type
Numeric
Numeric or Alpha or Graphic
Display Smoothing
Fast (real time)
Fast or Medium or Slow
Graph Time
Real Time
Real Time or 1 hour or 1 day or 1 week
Mute Button
Enabled
Disabled or Enabled
Horn
Enabled
Disabled or Enabled
Units
Imperial
Imperial or Metric
Digital Input 1
Disabled
Disabled or Enabled
Digital Input 1 Configuration
Disabled
External Alarm or Occ / Unocc mode
Digital Input 1 state
Disabled
Closed = Normal or Open = Normal
Digital Input 1 Occupied/Unoccupied
state
Disabled
Open = Occupied or Closed = Occupied
Digital Input 2
Disabled
Disabled or Enabled
Digital Input 2 Configuration
Disabled
External Alarm or Occ / Unocc mode
Digital Input 2 State
Disabled
Closed = Normal or Open = Normal
Digital Input 2 Occupied/Unoccupied
state
Disabled
Open = Occupied or Closed = Occupied
Password
00000
00000 to 99999
Re-beep mode
Disabled
Disabled or Enabled
Re-beep delay
Disabled
0 to 1000 seconds
Numeric mode range
1FPM
1 or 5 FPM
External Alarm 1 Delay
0 seconds
0 to 60 seconds
External Alarm 2 Delay
0 seconds
0 to 60 seconds
External Alarm 1
Disabled
Disabled or Enabled
External Alarm 2
Disabled
Disabled or Enabled
Relay Output 1
Disabled
Disabled or Enabled
Relay Output 1 Configuration
Disabled
Alarm or Purge
Relay Output 1 State
Disabled
Normally Open or Normally Closed
Relay Output 1 Purge State
Disabled
Normally Open or Normally Closed
Relay Output 2
Disabled
Disabled or Enabled
Relay Output 2 Configuration
Disabled
Alarm or Purge
Relay Output 2 State
Disabled
Normally Open or Normally Closed
Relay Output 2 Purge State
Disabled
Normally Open or Normally Closed