Configuration of the APM
Selection 6a
: Unoccupied Low Alarm Set Point
[SCREEN 6C]
By choosing unoccupied mode for the digital input, the APM
needs a new set of alarm points to “reset” to
during unoccupied mode. Press the
UNOCC LOW ALARM
button to set the new unoccupied low alarm
set point. By using the plus and minus buttons [
and
] the set point value can be adjusted. Press
ACCEPT
when the correct value is displayed, or press the
DISABLE
button to completely disable the low
alarm set point during unoccupied mode
[SCREEN 9C]
.
Selection 6b
: Unoccupied Low Alarm Delay Time
If an unoccupied low alarm set point was entered, the next screen displayed is the delay time selection
screen
[SCREEN 10C]
. By using the plus and minus buttons [
and
] the time delay, in seconds, can be
increased or decreased as desired. Press
ACCEPT
when completed. Note: a value of 0 turns off the
delay.
Selection 6c
: Unoccupied Low Alarm Caution Band
[SCREEN 11C]
If desired, a caution band can be utilized to warn the operator of an impending alarm condition. The
caution band is a value in FPM or MPS that is added to the unoccupied low alarm set point and will
display a warning message and flash the normal green LED until the unoccupied low alarm set point is
reached, then the unit goes into full alarm. For example if the unoccupied low alarm set point is set to
50FPM and the unoccupied low caution band is set to 10FPM; the APM will display a warning message
for all readings between 51 and 60FPM. Once the airflow velocity drops to 50 or below, the APM will go
into alarm mode. Readings above 60FPM are displayed in the normal fashion. By using the plus and
minus buttons [
and
] the caution band, in FPM or MPS, can be increased or decreased as desired.
Press
ACCEPT
when completed. Note: a value of 0 turns off the caution warning completely.
Selection 7a
: Unoccupied High Alarm Set Point
[SCREEN 12C]
Press the
UNOCC HI ALARM
button to set the new unoccupied high alarm set point. By using the plus
and minus buttons [
and
] the set point value can be adjusted. Press
ACCEPT
when the correct
value is displayed, or press the
DISABLE
button to completely disable the high alarm set point during
unoccupied mode.
Selection 7b
: Unoccupied High Alarm Delay Time
If an unoccupied high alarm set point was entered, the next screen displayed is the delay time selection
screen
[SCREEN 13C]
. By using the plus and minus buttons [
and
] the time delay, in seconds, can be
increased or decreased as desired. Press
ACCEPT
when completed. Note: a value of 0 turns off the
delay.
Selection 7c
: Unoccupied High Alarm Caution Band
[SCREEN 14C]
If desired, a caution band can be utilized to warn the operator of an impending alarm condition. The
caution band is a value in FPM or MPS that is subtracted from the unoccupied high alarm set point and
will display a warning message and flash the normal green LED until the unoccupied high alarm set point
is reached, then the unit goes into full alarm. For example if the unoccupied high alarm set point is set to
130FPM and the unoccupied high caution band is set to 25FPM; the APM will display a warning message
for all readings between 105 and 129FPM. Once the airflow velocity increases above 130FPM or higher,
the APM will go into alarm mode. Reading below 105FPM are displayed in the normal fashion. By using
the plus and minus buttons [
and
] the caution band, in FPM or MPS, can be increased or decreased
as desired. Press
ACCEPT
when completed. Note: a value of 0 turns off the caution warning completely.
Selection 8
: Exiting the Unoccupied Configuration Menu
When complete, press the
EXIT
button to go back to the
DIGITAL INPUTS
configuration options screen.
Press the
EXIT
button once again to go back to the
CONFIGURATION MENU
screen.