Routine Operation
Changing the heating block
58
PLEX-ID SP User Manual 10/2009 N09455
Abbott Laboratories
Table 4–4.
Processing volumes vs. plate types and magnet heads
Head / Plate
24 Deep Well plate
96 Deep Well plate
96 Well plate
Head for the 24
Deep Well plate
200 – 5000 µl
–
–
Head for
the 96
Deep Well plate
–
50 – 1000 µl
20 – 150 µl
Changing the heating block is controlled by the software. To change
the heating block, choose the Change_heatblock protocol by using the
Up
and
Down
arrow keys. Then press
START
. The protocol requests
you to remove the magnets and in the end to insert them. The heating
block will thus rise. Remove the heating block and then replace it
(Figure 4–44 and Figure 4–45). Insert the heating block by first setting
it and then pressing it hard into place. The heating block will snap at
both ends if you have done it correctly. Press the
START
button when
the task has been completed.
Figure 4–44.
Inserting the heating block
Changing the
heating block
Summary of Contents for PLEX-ID SP
Page 1: ...PLEX ID SP User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Page 6: ......
Page 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Page 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Page 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Page 92: ......
Page 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...