PLEX-ID SP User Manual 10/2009 N09455
39
Chapter 4
Routine Operation
This section on Do’s and Don’ts summarizes all the relevant procedures
on what to do and what not to do.
●
In case of any emergencies occurring during operation, switch
OFF and unplug the instrument immediately. Carry out
corrective measures. If the corrective measures taken do not
help, contact authorized technical service.
●
Carry out the operational check prior to normal use.
●
Only when the instrument is switched off, can you push the
heads or turn the turntable manually.
●
Always use correct volumes on the plates because the
movements of the tip comb are conducted according to the
volumes.
●
Use for self-testing is excluded.
●
Do not touch or loosen any screws or parts other than those
specifically designated in the instructions. Doing so might cause
misalignment and will invalidate the instrument warranty.
●
Never open any other cover of the PLEX-ID SP than the see-
through lid or the sliding door (Figure 2–3) while the
instrument is plugged into a power source.
●
Do not use the instrument if it appears that it does not function
properly.
●
Do not under any circumstances use formaldehyde.
●
Do not spill any alkalines onto any instrument surfaces to avoid
damage of the instrument. If needed, use suitable protection
covering.
Before you switch on the PLEX-ID SP, ensure that the voltage on the
type label at the bottom left of the back/side panel (Figure 2–4)
corresponds to the local voltage.
Warning
Never operate your instrument from a power outlet that has
no ground connection.
▲
Do’s and Don’ts of the
PLEX-ID SP
Do
Don’t
Switching on
Summary of Contents for PLEX-ID SP
Page 1: ...PLEX ID SP User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Page 6: ......
Page 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Page 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Page 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Page 92: ......
Page 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...