Installation
Installation setups
26
PLEX-ID SP User Manual 10/2009 N09455
Abbott Laboratories
To remove both the transport locks, follow these steps:
Figure 3–9.
PLEX-ID SP transport locks fitted
1.
To remove the transport lock of the tip comb holder, take firmly
hold of the lock and press tightly on both sides of the lock
according to Figure 3–10 so that the lock ends are released. Keep
hold of the tip comb holder and lock simultaneously, and start
gradually lifting the holder.
Figure 3–10.
Removing the transport lock of the tip comb holder
2.
Lift the tip comb holder higher and continue pressing the transport
lock of the tip comb holder. Remove the transport lock from its
location by ejecting it from the two eccentric pins (Figure 3–11).
When relocating the instrument, refer to “How to refit the
transport lock of the tip comb holder” on page 37.
Heating block
transport lock
Tip comb holder
transport lock
Summary of Contents for PLEX-ID SP
Page 1: ...PLEX ID SP User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Page 6: ......
Page 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Page 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Page 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Page 92: ......
Page 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...