Installation
Installation setups
28
PLEX-ID SP User Manual 10/2009 N09455
Abbott Laboratories
Note
Keep the transport locks (Figure 3–11 and Figure 3–13) for
future transportation of the instrument.
▲
Both the transport locks have been successfully removed. When
relocating the instrument, refer to “How to refit the transport lock
of the heating block” on page 36 and “How to refit the transport
lock of the tip comb holder” on page 37.
1.
Connect the mains supply cable to the mains power connector
(Figure 3–14) at the bottom left of the back/side panel. If you need
to use any other type of mains supply cable than supplied, use only
cables certified by the local authorities. Before you plug in the
power cable, ensure that the voltage on the type label at the bottom
left of the back/side panel (Figure 2–4) corresponds to the local
voltage.
2.
Connect the instrument to a correctly installed line power outlet
that has a protective conductor that is grounded.
Warning
Ensure that the mains switch (Figure 2–4) on the bottom left
of the back/side panel is in the OFF position. Never operate your
instrument from a power outlet that has no ground connection.
▲
Caution
Always connect or disconnect the serial cable when the power
is off.
▲
Figure 3–14.
Mains supply cable and serial cable connected
This section describes the installation setups of the heating block, shield
plate, interchangeable magnet heads and the see-through lid.
Fastening/changing the heating block is controlled by the software. To
fit the heating block, choose the Change_heatblock protocol under the
Maintenance
menu by using the
Up
and
Down
arrow keys. Then press
START
. The heating block will thus rise. Insert the heating block by first
setting it and then pressing it hard into place (Figure 3–15 and
Figure 3–16). The heating block will snap at both ends if you have
done it correctly.
How to ensure
startup
How to fit the subassemblies
of the instrument into place
Heating block
Summary of Contents for PLEX-ID SP
Page 1: ...PLEX ID SP User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Page 6: ......
Page 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Page 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Page 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Page 92: ......
Page 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...