Installation
How to pack the instrument for transportation
36
PLEX-ID SP User Manual 10/2009 N09455
Abbott Laboratories
This section describes how to refit both the transport locks, the
transport lock of the heating block and that of the tip comb holder.
To refit the transport lock of the heating block, follow these steps:
1.
Remove the heating block in the following way:
•
First choose the Change_heatblock protocol.
•
Then press
START
.
•
Remove the heating block by pulling it out.
•
Press the
START
button when the task has been completed.
2.
Switch the power OFF.
3.
Take the transport lock screw from its storage site.
4.
Insert the transport lock screw into the label and refit the transport
lock of the heating block into its place (Figure 3–25).
Figure 3–25.
Refitting the transport lock of the heating block (A)
5.
The transport lock of the heating block is refitted according to
Figure 3–26. Screw the transport lock screw clockwise
(Figure 3–26). Figure 3–29 shows both the transport locks refitted.
How to pack the
instrument for
transportation
How to refit the
transport lock of the
heating block
Summary of Contents for PLEX-ID SP
Page 1: ...PLEX ID SP User Manual Rev 1 0 ...
Page 2: ......
Page 3: ...PLEX ID SP User Manual Rev 1 0 Cat no N09455 ...
Page 6: ......
Page 78: ...Frequently Asked Questions Q As 76 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 86: ...Glossary 84 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories ...
Page 89: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 87 Notes ...
Page 90: ...88 PLEX ID SP User Manual 10 2009 N09455 Abbott Laboratories Notes ...
Page 91: ...Abbott Laboratories PLEX ID SP User Manual 10 2009 N09455 89 Notes ...
Page 92: ......
Page 93: ...Abbott Molecular Inc 1300 E Touhy Ave Des Plaines IL 60018 USA www abbottmolecular com N09455 ...