52 ID NOW Instrument User Manual
12.1.5 Run Test / Run QC Test Errors
Error: Timeout
The instrument has timed out during any of the steps in the testing process. More
than the defined amount of time elapsed during one of the steps outlined below.
For example:
1.
Between inserting the Test Base and inserting the Sample Receiver.
2.
Between the completion of the warm up and the addition and confirmation of
the sample
3.
Between adding the sample and transferring it from the Sample Receiver to the
Test Base.
4.
Between transferring the sample and closing the lid.
Action: Press OK. Remove and dispose of all test pieces and repeat with a new
test. Follow the on-screen instructions closely.
Error: Lid Open
The lid must not be opened during the testing process.
Action: Press OK. Remove and dispose of all test pieces and repeat with a new test.
Error: Test Failed
The test has encountered an unrecoverable error.
Action: Press OK. Remove and dispose of all test pieces and repeat with a new test.
Error: Invalid Frequency
If the User enters a non-numeric or numerical value greater than 1000, for the
QC frequency, this Error will appear.
Action: Press OK and enter a numeric value between 1 and 1000.
Error: QC Incomplete
For assays configured with QC Lockout set to
Warn or Lockout, this Error will
appear when a new patient test is attempted with a lot that does not have a valid
QC Positive and Negative result.
Action: Select OK to stop the patient test and proceed to conduct a QC run for
the lot OR the user must use a different lot of consumables for the run.