46 ID NOW Instrument User Manual
Step 7: Remove the USB drive. Turn the instrument off by removing the power
cable from the instrument. (Do not use the power button to power off the
instrument.) Wait 10 seconds and then reattach the power cable to the
instrument. Power up the ID NOW Instrument by pressing and holding
the power button. The instrument will power up as normal.
Prior to running an assay with the new software for the first time, the user is
required to run a positive and negative successful QC test before patient testing is
allowed. See Section 6.
If expected QC results are not obtained or you are unable to complete the software
update, please contact Abbott Technical Support.
Note: Only one USB drive may be present during the software update process.