ID NOW Instrument User Manual 17
4 SETUP MENU
Home>Setup
Setup
Login
Users
Assay Preferences
ID & Password Required
Patient Details
99
The
Setup screen allows the Admin to configure the
instrument. Use the up and down arrows to scroll
through menus.
Settings are saved even when power is off.
Settings include:
• Login
• Users
• Assay Preferences
• Patient Details
• Date
• Time
• Auto-logout
• Network
• Remote Mode
• Connectivity
• QC Lockout
• Lot Management
• Auto-Print
• Reset to Factory Default
Chosen options are displayed to the right of each
setting.
4.1 Log In (Admin Access Only)
Home>Setup>Login
Log In
User ID Only
OK
User ID & Password
Log In allows the Admin to choose User ID
password requirements to allow access to the
instrument.
If Users do not require passwords, select ‘User ID
Only’ and press OK. If a password is preferred,
select ‘User ID & Password’ and press OK.
The selected option will apply to all Users with the exception of the Admin, who will
always be required to use a password when logging on as ‘Admin’.
If the system is changed from User ID & Password to User ID Only, existing passwords
are retained in the event the Admin selects User ID & Password in the future.
If the system is changed from User ID Only to User ID & Password, user passwords are
set to ‘password’ as a default, if a password has not been set previously for the user. The
Admin can change each User’s password individually, as needed.