ID NOW Instrument User Manual 9
1.6 Maintenance & Cleaning
The ID NOW Instrument is maintenance-free and has no serviceable parts. In the
case of instrument failure or damage, contact Abbott Technical Support.
Acceptable cleaning agents include the following:
• 70% ethanol - available in commercial wipes or on a damp, lint free cloth
• 70% isopropanol - available in commercial wipes or on a damp, lint free cloth
• 10% bleach - on a damp, lint free cloth only
Do not spray or pour solution directly onto instrument when cleaning. Ensure no
excess liquid is used when cleaning as it may damage the instrument.
Abbott recommends that the exterior instrument surfaces and the surfaces visible
under the open lid be cleaned daily. Clean surrounding bench area. Clean instrument
and surrounding areas immediately after possible patient sample contamination.
Do not disassemble the instrument for cleaning.
Do not immerse in water or cleaning solutions.
Do not clean with soap or other solutions.
1.7 Components
The following items are supplied with the ID NOW Instrument:
• Power Supply and Adaptor
• ID NOW Instrument User Manual
• ID NOW Instrument Quick Start Guide