4-9
2000-OSM, F1
Parameters) soft key. On the Control Parameters screen (see Figure 4-9), cursor down the
selections until "Detector 1 Mode" and "Detector 2 Mode" appear. Select the desired mode for
Detector 2 and then press the F2 (Exit and Update) soft key to save this setting.
**********
Control Parameters
**********
Update Trends on Alarm:
No Yes
Auto Update of RF:
No
Yes
Max RF Variation:
0.000000
Unknown Peak RF:
0.000000
Unknown Peak Units:
%
ppm
Slope Detect Tics: Never Last First
Detector 1 Mode: Master
Detector 2 Mode: Master Slave
Analyzer Name:
More Below
Exit
Alarm
Common
Escape and
Mask
Digital
Update
Out. Mask
F1
F2
F3
F4
Figure 4-9. CONTROL PARAMETERS SCREEN
The Stream Assignment screen adds two columns to its standard screen (see Figure 4-10), one to
identify the detector used for each stream and method (Det), the other to identify the link between
master and slave (Link).
******
Stream Assignment
******
Stream
Status Method Basic Det Link
01
Active
01
00
01
02
02
Active
02
00
02
00
03
Active
01
00
01
04
04
Active
02
00
02
00
05
Inactive
01
00
01
00
06
Inactive
01
00
01
00
07
Inactive
01
00
01
00
08
Inactive
01
00
01
00
Exit
F1
F2
F3
F4
Figure 4-10. STREAM ASSIGNMENT SCREEN