2000-OSM, F1
7-29
CALIBRATION STREAM METHOD TABLE
Method
Table
#10
Time
Function Value
0001
Temp.
Contrl
One
0120
Noise Calc On
0125
Noise
Calc
Off
0128
Auto
Zero
N/A
0132
Valve
On
One
0155
Valve
Off
One
0160
Slope Detect
On Now
0160
Peak
Width
Eight
0175
Component One
0175
Component One
0182
Component Two
0188
Component
Three
0208
Component
Four
0217
Component Five
0217
Baseline
Force
@
Valley
0247
Component Six
0262
Component
Seven
0290
Component
Eight
0290
Baseline
Force
@
Valley
0312
Component
Nine
0312
Baseline
Force
@
Valley
0343
Component Ten
0343
Baseline
Force
@
Valley
0359
Component
Eleven
0359
Baseline
Force
@
Valley
0390
Component
Twelve
0390
Baseline
Force
@
Valley
0424
Component
Thirteen
0424
Baseline
Force
@
Valley
0440
Component
Fourteen
0440
Baseline
Force
@
Valley
0462
Component
Fifteen
0462
Baseline
Force
@
Valley
0509
Component
Sixteen
0509
Baseline
Force
@
Valley
0540
Component
Seventeen
0540
Baseline
Force
@
Valley
0570
Component
Eighteen
0570
Baseline
Force
@
Valley
0601
Component
Nineteen
0601
Baseline
Force
@
Valley
0650
Slope
Detect Off
0710
Stream
Step N/A
The various time coded functions are defined as follows. Section 4 contains more complete
descriptions of these functions.
0001 Temp.
Contrl One
This function ensures that the temperature of the isothermal oven is at the set point value prior to
sample injection.