![ABB 2600T Pressure Transmitter Series User Manual Download Page 25](http://html.mh-extra.com/html/abb/2600t-pressure-transmitter-series/2600t-pressure-transmitter-series_user-manual_3724025.webp)
364xS User’s Manual
25
Configuration
Set the Damping configuration by PC
Damping configuration by PC is available using dedicated Software. ABB can provide the references and the
Software. Please refer to the Software instructions for the appropriate operations.
Set the Units
To monitor the process using the appropriate units follow one of the methods described in the next
paragraphs.
Set the Units by Digital LCD Integral Display
To change the unit in the Digital LCD Integral Display
Navigate the Digital LCD Integral Display menu, select
Display
>
Display Settings
>
Display Eng. Unit
, and
the possible Engineering variables will be available. To select the type, push
Edit
and scroll with a up and
down keys; confirm with
OK.
Pressure
Temperature
Flow
Mass & Volume
Density
Level
Other
Custom Unit
The in-use unit for the type will be available. To modify select
Edit
then scroll the desired unit (see the
following tables) with the up and down keys and confirm with
OK.
Units for Pressure variable
mbar
millibar
mmH2O°C
millimeter of water at 4 degrees Celsius
g/cm
2
grams per square centimeter
inH2O°F
inches of water at 68 degrees Fahrenheit (20°C)
Kg/cm
2
kilo grams per square centimeter
inHg°C
inches of mercury at 0 degrees Celsius
Pa
pascal
ftH2O°F
feet of water at 68 degrees Fahrenheit (20°C)
kPa
kilopascal
mmH2O°F
millimeter of water at 68 degrees Fahrenheit (20°C)
torr
torr
mmHg°C
millimeter of mercury at 0 degrees Celsius
atm
atmosphere
psi
pounds per square inch
MPa
Megapascal
bar
bars
inH2O°C
inches of water at 4 degrees Celsius
Units for Temperature variable
°C
degree Celsius
°F
degree Fahrenheit
°R
degree Rankine
K
Kelvin
Summary of Contents for 2600T Pressure Transmitter Series
Page 1: ...InstrumentationIT 2600T Pressure Transmitter family ABB User s Manual 364DS and 364PS Models ...
Page 7: ...364xS User s Manual 7 This page has been intentionally left blank ...