15-3-3 DAV/DAI voltage/current check
This mode allows the user to check the DAV/DAI value. To use this mode, first enter the check mode.
Voltage is used for checking. To check the current value (DAI), the displayed value of 0 corresponds
to 4 mA and the value of 10 corresponds to 20 mA. No negative values are available for the DAI
values and the increase of the displayed value by one corresponds to the increase of the current
value by 1.6 mA.
This mode will be displayed only when the DAV/DAI option is installed.
1. In the measurement mode, press the FNC key for three seconds or
more to enter the function selection mode.
2. Press the > key several times to display
0p-da
.
3. Press the
↵
key to enter the DAV/DAI check mode.
1
>
Moves to the display check mode without performing the
DAV/DAI check mode.
ESC Cancels the operation and returns to the
CHeCk
1
display.
4. In the main display, the DAV output voltage is displayed, Connect the
digital multimeter. Check the displayed value and output voltage
value.
To check the current value (DAI), the displayed value of 0
corresponds to 4 mA and the value of 10 corresponds to 20 mA. The
increase of the displayed value by one corresponds to the increase
of the current value by 1.6 mA.
1
∧
Increases the output voltage value by one. After 10 V, 0
V appears. (Increases the output current value by 1.6.)
1
∨
Decreases the output voltage value by one. After -10 V, 0
V appears. (Decreases the output current value by 1.6.)
1
↵
Finishes the DAV/DAI check mode and moves to the
display check mode.
ESC
Cancels the operation and returns to the
0p-da
1
display.
FNC
3 seconds
4532
1000 400
fnc
0p-da
0
CHeCk 0p-da
di5p
DAV: 0 V DAI: 4 mA
1
CHeCk 0p-da
DAV: 1 V DAI: 5.6 mA
10
CHeCk 0p-da
DAV: 10 V DAI: 20 mA
DAV: 10 V DAI: 20 mA
10
CHeCk 0p-da