5. REAR PANEL
This chapter explains the terminals on the rear panel and how to connect sensors.
For the position of each terminal, see the illustration below.
Terminal numbers that are printed on the top of
the indicator casing
Confirm the terminal numbers when making connections
When making connections, confirm the terminal numbers printed on the side of the
terminal block and on the top of the indicator casing.
5-1 Description of Each Terminal
5-1-1 Ground terminal (27) and AC input terminals (35) (36)
(27) Ground terminal
Be sure to connect a grounding wire to avoid electrical shock and prevent static electricity from
affecting measurements.
(35) (36) AC power input terminals
Connects the AC power cord. The power requirement is 85 VAC to 250 VAC, 50/60 Hz.
When making connections
Switch off the power of all the instruments used.
Keep cables sensitive to electrical noise away from power cables and other
sources of electrical noise.
Connect the ground terminal to earth ground with a relatively heavy cable to protect
the terminal circuits against surges.